Factors Affecting Polyurea Elongation Characteristics:
Product Specifications Vs Field Results

 
Lee Bower
President, PCSI

Dudley Primeaux
Primeaux Associates LLC
 

BACK

 

Background:

Since July, 2000, Polyurea Coating Systems, Inc. (PCSI) has been supplying the epoxy primer (PCS-30) and polyurea (PCS-355) to the applicator for the Remote Delivery Facility (RDF) waterproofing project. (please see “Pentagon Remote Delivery Facility “ case study). PCSI was asked in late June, 2000 to evaluate the RDF waterproofing project following the application of a coating system which was not performing as required. Numerous problems with the current system were evident; multiple pin hole formations, poor adhesion to the concrete, poor intercoat adhesion, and markedly lower performance characteristics than reported in the manufacturers product literature. After reviewing the project status, the applied coating, and the concrete surface, several suggestions were provided by PCSI technicians to the applicator and the general contractor on methods to provide a continuous waterproofing membrane on the deck of the RDF. 

PCSI is a manufacturer and supplier of polyurea and epoxy systems for coating concrete,  steel and other substates. Several products are offered, depending on application requirements, end use, and service conditions. PCSI has developed products based on a systems approach to an application. PCSI primers are matched with the PCSI polyurea to achieve optimum results of the finished coating. 

After reviewing the conditions presented at the RDF, PCSI offered a primer-polyurea system, which when properly applied, would meet the project requirements; waterproofing over 280,000 ft2 of cast concrete.  PCS-30, a two component 100% solids penetrating primer was chosen due to the varying conditions of the concrete. PCS-355 polyurea was applied at 80 mils minimum thickness over the primed deck. PCS-355 is reported on PCSI data sheets as having a typical tensile strength of 2,500-3,000 psi, elongation of 500%, and a hardness of 90-96 shore A. These test results are based on testing that was performed at the PCSI laboratories by trained laboratory technicians. Results are reported as typical of the specific formulation and are meant as a guideline when choosing a product for a particular application. 

Over 15,000 gallons of PCSI product has been applied to the RDF deck. PCSI has supplied over 400 man hours of on-site technical assistance to the applicator and the general contractor during the application, provided recommendations on repair of inferior concrete surface, and consulted on additional surfaces requested to be coated. 

Discussion: 

As a requirement of this project, Quality Control reports were generated for each lot of PCS-355 manufactured and sent to the RDF, with the testing performed in the PCSI analytical laboratories. Each QC report was submitted to the applicator and general contractor prior to shipment of the lot. Each lot was analyzed for tensile, elongation and hardness. The individual QC reports provide the average value of a minimum of 3 tests for each specimen tested. 

All lots of PCS-355 supplied to the RDF have met and exceeded the specifications established by PCSI and reported on PCSI’s technical data sheets. 

However, a concern was raised that, after cutting a section of the applied coating from the deck and sending it to an independent testing lab for analysis, the elongation values reported by PCSI were not confirmed by the testing lab. Elongation values determined by the testing lab showed 50-70% lower values than those reported by the PCSI labs. 

As mentioned above, PCSI performs QC analysis on each lot of PCS-355 prepared for the RDF. Each lot tested exceeded specification. In the event a lot did not meet PCSI’s specification, the material would be removed from shipping status. No lots of PCS-355 were manufactured outside of PCSI’s specifications. 

In addition, sample coupons were prepared for each lot sent to the RDF and retained in PCSI’s files for future testing. These retained coupons were re-analyzed following the questions raised by the contractor. Each coupon again met specification. Full backup data for the QC summary reports were provided to the general contractor. 

Therefore, the question remained, repeated analysis of the PCS-355 by PCSI laboratories were not reproduced by samples taken from the RDF deck and analyzed by an independent lab. Specifically the elongation values were 10-50 points lower than that reported on the PCSI specification data sheet for the product. 

What conditions could affect these varying test results? PCSI technicians were convinced that the testing apparatus was properly calibrated and tests were performed following the ASTM D 412 and ASTM D 638 standards. Reproducibility of the various lots and retains were within ASTM standard deviations, as reported in the Precision and Bias sections. Yet the product sprayed on the RDF deck, removed and tested, did not match our results. PCSI began to look at how application, and atmospheric conditions could affect the results. 

First, in order to support PCSI numerical data, samples of each lot of PCS-355 were again pulled in the tensile state using a model 1101 ATS machine with a 1,000 lb load cell. Prior to performing the test, each coupon was marked along the throat of the prepared specimen (or “dog bone”), ˝” either side of center.  A ruler was placed along the side of the specimen as the test began. After the coupon was stretched to over 500%, a photograph was taken visually demonstrating the elongation. Figure 1 shows an example of PCS-355, lot number 636JO. Aligning the marks along the ruler, the sample was stretched over 5 X’s it’s initial length, or >500%. Note that none of the samples in the photographs were broken, indicating that the elongation exceeded the values visually presented. All lots sent to the RDF were tested in this manner.
 

Figure 1 

ASTM D 638, Tensile Properties of Plastics, references note 2, describes the expected variations in test results between testing labs. “ This test method is not intended to cover precise physical conditions. It is recognized that the constant rate of cross-head-movement type of test leaves much to be desired from a theoretical standpoint, that wide differences may exist between rate of crosshead movement and rate of strain between gage marks on the specimen, and that the testing speeds specified disguise important effects characteristics of materials in the plastic state.” Note 2 goes on to say, “ special additional tests should be used where more precise physical data are needed.”  

Mr. Dudley Primeaux , consultant, recognized industry expert in polyurea chemistry and applications, and co-author of this study presented additional data demonstrating the non-precise nature of testing elongation of polyurea. In a documented test study, Mr. Primeaux provided the leading manufacturer of plural component spray equipment, Gusmer, with a master batch of polyurea. Gusmer sprayed this master batch with 8 different Gusmer gun models, holding all other application conditions constant. Panels were then sent for analysis. The results are remarkable, with elongations ranging from 96% to 248%, depending on the model and set up of Gusmer gun alone. The polyurea used was a 250% elongation polyurea per spec data sheet. All other conditions were held constant. Results from this study are shown in Table 1. 

 
 
Table 1
Spray Gun Elongation, %
(a) 452 w/212 PCD, 2500psi 216
(b) GX-7 new block, 402 w/55 PCD, 2 passes 183
(c) Same as above, 1 pass 149
(d) GX-7 new block, impinger, 2 passes 193
(e) GX-7 #15 Mod w/212 PCD mult. passes 248
(f) GX-7 #15 Mod w/212 PCD 1 pass 183
(g) Same as above 242
(h) GX-8 201 w/fan 214
(i) GX-8 202 w/fan 96
(j) GAP gun, 2 passes 241
(k) GAP gun, 1 pass 172
(l) Probler gun, .01 tip, 1 pass 215
   
   
 

Note the wide variation in %elongation in table 1. Results ranged from 96% elongation (i) to 248% (e). Similar gun set ups showed differences in elongation of 65 units by decreasing the passes of the spray, (e) and (f). Gun set up (g) is a repeat of (f), yet showed a 59 unit difference between the two. Choice of gun and set up of that gun will affect the results of elongation tests. 

Mr. Primeaux also referenced one of his papers, “A Study of Polyurea Spray Elastomer Systems” presented at the High Solids Coatings Conference in 1995. In this paper, analysis of processing temperature vs elastomer physical properties was presented. In this evaluation, the effect of the temperature of the polymer at the gun tip was measured against the elongation values of the resulting system. Table 2 presents the results from Mr. Primeaux’s  study. 
 

 

Table 2 

Mix Temperature, °F                100      129      140      151      160

 Elongation, %                           16        68        76        126      150

   
 

This clearly demonstrates that as the mix temperatures at the gun are lowered, elongation is similarly affected. Temperature conditions at the RDF site have been dropping as fall and winter approached. The applicator employs 300’ of spray hose between the heater and the gun tip. Although the heating block at the proportioner is set to specified temperatures and the hose is traced, the loss of heat through the hose while laying on the cold concrete deck will decrease the tip temperature and affect elongation values. 

Processing pressure will also affect elongation values. The most obvious characteristic of polyureas is the very rapid reaction between the isocyanate and the primary amines. As a result an impingement mix is required for processing polyurea. Temperature as well as pressure plays an important role in balancing viscosity, reactivity and mixing of the two components. As seen in Table 3, tip temperatures can be set within the limits of the equipment. Likewise, processing pressure can also be adjusted. Table 3 shows the affect of varying processing pressure while holding all other parameters constant. It is also imperative that the pressure gauges at the proportioner be constantly monitored and that the sprayer be notified as soon as drop in pressure or pressure differentials between A and B streams occur. 
 

 

 Table 3 

                     Mix pressure, psi          900      1000    1100    1400    1800    2000

                     Elongation %                14.4     40.1     71.5     87.8     158      151

Mix temperature of 71°C. 
 

  Sample thickness has an affect on elongation values as well. Mr. Primeaux again references work performed in 1996, which compares elongation against sample thickness. All other parameters were constant. Table 4 shows that as sample thickness is increased, elongation decreases. 
 

Table 4 

Avg Thickness, in               .027     .066     .085

 Elongation, %                     390      275      280

 

 

Sample thickness will change throughout the area being sprayed due to wind conditions, and applicator techniques. Lab evaluations are made at 60 mil sheets. Tests at the RDF showed film thickness varying from 70 mils to over 100 mils. Elongation values will reflect the differences in coating thickness. 

Cure time of the product will also affect elongation values. Although polyurea sets up and is dry to the touch in seconds, some systems take longer to develop full physical properties. Graph 1 shows the development of elongation over time of PCS- 355.
 

 

Graph 1

                         

 

Polyurea formulations differ between products and manufacturer. In the case of PCS-355, ultimate elongation values of 500% are reached after one week of cure time. If samples were obtained and analyzed in less than a weeks time, differences in elongation will be observed. 

Summary

Typically, manufacturers of polyurea products report performance data of the supplied material in their technical data sheets. This data may consist of parameters such as tensile strength, hardness, and elongation. Additional tests may be supplied such as abrasion resistance, coefficient of thermal expansion, flexibility etc. These tests may be performed in house or by outside laboratories depending on the availability of expensive testing apparatus. Coupons prepared for analysis are made under laboratory conditions and controlled environments. The resulting data is designed to provide an indication of the performance values of the formulation. However, care must be taken in transferring these lab produced results to actual field applications. This reports demonstrates how application variables and conditions may result in a wide range of results performed on the same polyurea formulation. Simply utilizing a different brand or model spray gun, or using a different temperature or pressure at the gun tip may results in variations from reported values. Sample thickness and cure time also influence results for elongation. Atmospheric conditions at the application site will change day by day and will affect property development. Finally, it is recommended that the ASTM procedures be carefully read for expected deviations in data reported between laboratories.

[TOP]

 

BACK

 

Copyright 2002 Polyurea Coating Systems, Inc.
PCSI - (281) 565-7001